How To Change The Display Settings
You can read the steps below to see how you can change the display settings. If you want to print a guide with images to keep beside you while performing the actions, you can download this file.

Changing Settings Of The LCD 23 Display
Step 1
Turn OFF the leak tester.
Step 2
Turn ON the leak tester while pressing the Enter button.
Step 3
LCD Parameter will appear. After releasing the enter button the first setting ( Display no.) will appear.
With the arrow buttons you can scroll though the menu:
Set: display no.
Set: OEM
Set: end.
To enter one of the sub menus press enter. In the sub menu scroll with the arrow buttons to the correct setting and press enter to select.
Oxipack setting are:
Display no. : No. 0 single
OEM: Jetter
When correct settings are selected go to sub menu Set: end and select with the Enter. The display will reboot and settings are stored.
Display no. : No. 0 single
OEM: Jetter
When correct settings are selected go to sub menu Set: end and select with the enter. The display will reboot and settings are stored.