Procedure Before Each Shift
Read the steps below to understand the procedure to perform before starting testing at each shift. You can also download this document if you want to print this information.

Procedure Before Each Shift
Step 1
Clean the membranes and seal with a damp cloth or alcohol wipe.
Step 2
Perform an empty machine test.
Result should always be: ‘Pack OK’.
With a ‘Pack OK’ result, you have successfully checked the integrity of the Oxipack leak tester.
In case you have a different result:
- Error 202/201/101
The machine has a leak; check for damages on the membrane and/or seal. If damages are found on the membrane or seal: replace the membrane or replace the seal. If debris or dirt is found on the seal, clean the seal with a damp cloth or alcohol wipe. - Error 203
Non-return valve is leaking: contact Oxipack support immediately. - Perform a dummy test
Result should always be: ‘Pack leak’. With a Pack leak result you have successfully checked the ability of the Oxipack leak tester to find micro leaks.
Step 3
Perform a dummy test.
Result should always be: ‘Pack leak’.
With a Pack leak result you have successfully checked the ability of the Oxipack leak tester to find micro leaks.
In case you have the result: ‘Pack OK’
The micro-hole in the insert is clogged or blocked: contact Oxipack support immediately.
Step 4
The Oxipack leak tester is ready to test your packages.